Description: This antibody has been selected for its ability to recognize rmMMP-9 in direct ELISAs and western blots in which this antibody shows approximately 10% cross reactivity with rhMMP-9 and less than 2% cross reactivity with rhMMP-1, -2, -3, -7, -8, -10, -12 and -13. This antibody is suitable for immunohistochemistry.
Primary Antibody
|
Name: MMP-9 Antibody |
|
Clone: Goat anti-Mouse |
|
Supplier: R&D Systems |
|
Catalog Number: AF-909 |
|
Dilution: 1:50 using IHC-TekTM Antibody Diluent (Cat# IW-1000 or IW-1001) to reduce background and unspecific staining and serum blocking step is NOT needed. |
|
Incubation Time/Temp: 60 min/room temperature |
| Device: IHC-TekTM Epitope Retrieval Steamer Set (Cat# IW-1102) |
| Buffer/pH value: IHC-TekTM Epitope Retrieval Solution (Cat# IW-1100) |
| Heat/Cool Temperature: 95-100 ºC/room temperature |
| Heat/Cool Time: 20 minutes/20 minutes |
| Standard Method: ABC Method or LSAB Method |
| Enhanced Method: Polymeric Methods |
| Reagent: DAB |
| Incubation Time/Temperature: 1-3 minutes/room temperature |
| Reagent: Gill's Hematoxylin or Mayer's Hematoxylin |
| Staining Time: 30 seconds |
| Staining Pattern: Cytoplasmic/nuclear |
| Images: Search image |
| Tissue Type: Lung |
| Fixation: Formalin fixed paraffin sections |
| Positive Control: Lung |
| Negative Control: Omit primary antibody, isotype control, absorption control |
| Blocking: 2-5% normal serum to reduce unspecific background staining; 0.5-3% H2O2 to block endogenous peroxidase activity; avidin/biotin to block endogenous biotin activity if necessary |